KWRRI Research Reports

Archived

This content is available here strictly for research, reference, and/or recordkeeping and as such it may not be fully accessible. If you work or study at University of Kentucky and would like to request an accessible version, please use the SensusAccess Document Converter.

Abstract

Construction and testing of a double-beam absorption spectrographic analysis system using a d. c. arc multielement source has been conducted. An optical system design which brings analytical and reference beams together to illuminate the upper and lower portions of the spectrograph slit has been shown to be functional. However, ad. c. arc will not serve as a multielement source for the intended purposes because of excessive thermal broadening of emission lines.

A direct excitation d. c. arc method was reviewed which can be used for the rapid determination of some cations in water.

Publication Date

10-1972

Issue Number

53

Digital Object Identifier (DOI)

https://doi.org/10.13023/kwrri.rr.53

Funding Information

The work on which this report is based was supported in part by funds provided by the Office of Water Resources Research, United States Department of the Interior, as authorized under the Water Resources Research Act of 1964.